Entry  Tue Jun 14 18:45:43 2022, Anchal, Update, SUS, LO2 lower magnets are stuck in coil, won't come off 
    Reply  Tue Jun 14 19:34:06 2022, yuta, Update, SUS, Resonant frequency identification from the free swing test 7x
       Reply  Thu Jun 16 18:30:07 2022, Anchal, Update, SUS, New diagonalized input matrices applied SUSInpMatDiag_MC1_MC2_MC3_BS_IMTX_ETMX_ETMY_AS1_AS4_SR2_PR2_PR3.pdf
          Reply  Tue Jun 21 08:36:50 2022, Anchal, Update, SUS, Diagonalized input matrices for LO1, freeSwing on ITMY and ITMX SUSInpMatDiag_LO1.pdf
    Reply  Wed Jun 15 15:07:07 2022, Koji, Update, SUS, LO2 SUS stuck fixed Screenshot_2022-06-15_15-00-22.pngScreenshot_2022-06-15_15-00-40.png
Message ID: 16918     Entry time: Wed Jun 15 15:07:07 2022     In reply to: 16913
Author: Koji 
Type: Update 
Category: SUS 
Subject: LO2 SUS stuck fixed 

I checked the state of the LO2 suspension. I found that the coil driver Enable Mon was all red. Meaning, the actuation signals were not delivered to the coil driver. I wasn't sure if this was intentional or not.

Enabled the coils with "WD Master" Shutdown -> Normal.

Immediately I saw the OSEMS flipped the sign because there was an (non-intentional) alignment offset in pitch. I've adjusted the pitch offset so that all the OSEM PDs have the voltages 4~5V.

That's it.

Attachment 1: Screenshot_2022-06-15_15-00-22.png  33 kB  Uploaded Wed Jun 15 16:07:45 2022  | Hide | Hide all
Screenshot_2022-06-15_15-00-22.png
Attachment 2: Screenshot_2022-06-15_15-00-40.png  13 kB  Uploaded Wed Jun 15 16:07:52 2022  | Hide | Hide all
Screenshot_2022-06-15_15-00-40.png