Tue May 24 11:04:53 2022, JC, Update, VAC, Beginning Pumpdown 
|
Wed May 25 16:38:27 2022, yuta, Update, SUS, Suspensions quick health check 
|
Tue Jun 7 17:26:21 2022, yuta, Update, SUS, ITMY ULCOIL mystery not solved  
|
Tue Jun 7 19:13:12 2022, yuta, Update, SUS, Agreement on suspension damping loop polarity conventions
|
Fri Jun 10 15:02:04 2022, yuta, Update, SUS, Servo gain sign flipped for MC1 WFS relief
|
Tue Jun 7 19:40:45 2022, Anchal, Update, SUS, ITMY changed output matrix to disable use of UL coil 
|
Wed Jun 8 18:16:20 2022, yuta, Update, SUS, ITMY ULCOIL mystery: Coil driver swap test 
|
Thu Jun 9 23:08:39 2022, rana, Update, SUS, ITMY ULCOIL mystery: Coil driver swap test
|
Fri Jun 10 13:02:14 2022, yuta, Update, SUS, ITMY ULCOIL mystery: Coil driver swap test
|
Fri Jun 10 15:04:23 2022, rana, Update, SUS, ITMY ULCOIL mystery: Coil driver swap test
|
Fri Jun 10 21:10:01 2022, yuta, Update, SUS, ITMY ULCOIL mystery: Coil driver swap test
|
Wed Jun 15 17:03:17 2022, yuta, Update, SUS, ITMY ULCOIL issue solved, loose connection in sat amp box   
|
Wed Jun 15 15:45:37 2022, yuta, Update, SUS, LO1 LLCOIL now working, it was loose connection  
|
Wed Jun 15 07:26:35 2022, JC, Update, VAC, Beginning Pumpdown
|
|
Message ID: 16904
Entry time: Thu Jun 9 23:08:39 2022
In reply to: 16903
Reply to this: 16905
|
Author: |
rana |
Type: |
Update |
Category: |
SUS |
Subject: |
ITMY ULCOIL mystery: Coil driver swap test |
|
|
what was the result of the inductance measurement? should be ~3.3 mH as measured from the flannge or cable that goes to the flange from sat amp.
|