Message ID: 16876     Entry time: Thu May 26 15:55:10 2022
Author: Tommy 
Type: Update 
Category: Electronics 
Subject: RFSoC Power Spectrum 

Finished building power spectrum analyzer for the RFSoC. There are two things that I would like to address down the road. First is that there is an oscillation between positive and negative voltages at the ADC sampling frequency. This creates an undesirable frequency component at the sampling rate. I have not yet figured out the cause of this positive to negative oscillation and have simply removed half of the samples in order to recover the frequency. Therefore, I would like to figure out the root of this oscillation and remove it. Also, we have a decimation factor of 2 as default by the board which we would like to remove but have been unable to do so.

Example: 8 MHz Square Wave from SRL signal generator.

Attachment 1: Screen_Shot_2022-06-02_at_2.51.23_PM.png  140 kB  Uploaded Thu Jun 2 15:52:37 2022  | Hide | Hide all
Screen_Shot_2022-06-02_at_2.51.23_PM.png