Message ID: 16778     Entry time: Thu Apr 14 10:18:35 2022     In reply to: 16635     Reply to this: 16780
Author: Paco 
Type: Summary 
Category: BHD 
Subject: 2 in oplev mirrors incompatible with LMR2V 

[Paco, JC]

We realized the 2 in oplev mirrors (Thorlabs BB2-E02) for ITMYOL, SRMOL, and BSOL, are 0.47 in thick, while the LMR2V fixed mount is 0.46 in deep, even without taking the retaining ring into account. After a brief exchange with Koji, and Ian, we decided to glue the mirrors onto the mounts using Torr Seal (a vac compatible epoxy). They are curating in the clean room and should be ready to install in about 2 hours.