Sat Feb 26 01:01:40 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels
|
Sat Feb 26 15:45:14 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels
|
Mon Feb 28 19:15:10 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels
|
Tue Mar 1 20:38:49 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels   
|
Wed Mar 2 11:50:39 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels    
|
Wed Mar 2 12:40:08 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels 6x
|
Sun Mar 6 18:14:45 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels
|
Mon Mar 7 13:53:40 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels
|
Mon Mar 7 14:52:34 2022, Koji, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels
|
Tue Mar 8 12:08:47 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels
|
|
Message ID: 16713
Entry time: Tue Mar 8 12:08:47 2022
In reply to: 16707
|
Author: |
Tega |
Type: |
Update |
Category: |
VAC |
Subject: |
Ongoing work to get the FRG gauges readouts to EPICs channels |
|
|
OMG, it worked! It was indeed a calibration issue and all I had to do was press the "OK" button after selecting the "CAL" tab beside the pressure reading. Wow.
Quote: |
Great trouble shoot!
> I guess, the only remaining issue now is the incorrect atmospheric pressure reading of 1000 Torrs.
This is just a calibration issue. The controller should have the calibration function.
(The other Pirani showing 850Torr was also a calibration issue although I didn't bother to correct it. I think the pirani's typically has large distribution of the calibration values and requires individual calibration)
|
|
|
|