We expect that the MC sus are susceptible to the temperature change and the alignment drifts away with time.
Here is the proper alignment procedure.
0) Assume there is no TEM00 flash or locking, but the IMC is still flashing with higher-order modes.
1) Use the CCD camera and WFS DC spots to bring the beam to the nominal position.
2) Use only MC2 and MC3 to align the cavity to have low-order modes (TEM00,01,02 etc)
3) You should be able to lock the cavity on one of these modes. Minimize the reflection (maximize the transmission) for that mode.
4) This should allow you to jump to a better lower-order mode. Continue alignment optimization only with MC2/3 until you get TEM00.
5) Optimize the TEM00 alignment only with MC2/3
6) Look at the MC end QPD. use one of the scripts in scripts/MC/moveMC2 . Note that the spot moves opposite to the name of the scripts. i.e. MC2_spot_down moves the spot up, MC2_spot_right moved the spot left, etc...
These scripts move MC1/2/3 and try to keep the good MC transmission.
7) moveMC2 scripts are not perfect. As you use them, it makes the MC alignment gradually degraded. Use MC2 and MC3 to recover good transmission.
8) If MC2 spot is satisfactory, you are done.
-------------
Step 6-8 can be done with the WFS on. This way, you can skip step 7 as the WFS servo takes care of it. But if the spot move is too fast, the servo can't keep up with the change. If so, you have to wait for the settling of the servo. Once the spot position is satisfactory, MC servo relief should be run so that the servo offset (in actuation) can be offloaded to the bias slider.
|