Entry  Sat Feb 26 01:01:40 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels 
    Reply  Sat Feb 26 15:45:14 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels Vac-gauges-flow-diagram.png
       Reply  Mon Feb 28 19:15:10 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels 
       Reply  Tue Mar 1 20:38:49 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels Working_Serial_Port_List_1.pngWorking_Serial_Port_List_2.pngWorking_Config_Ports#1-5.pngWorking_Config_Ports#7-8.png
          Reply  Wed Mar 2 11:50:39 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels XGS600_IOLAN_settings_1.pngXGS600_IOLAN_settings_2.pngXGS600_USBserial_settings.pngXGS600_comm_test.pyXGS600_comm_test_result.txt
             Reply  Wed Mar 2 12:40:08 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels 6x
                Reply  Sun Mar 6 18:14:45 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels iolan_xgs_comm_investigation.pdf
                   Reply  Mon Mar 7 13:53:40 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels iolan_xgs_comm_live.pdf
                      Reply  Mon Mar 7 14:52:34 2022, Koji, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels 
                         Reply  Tue Mar 8 12:08:47 2022, Tega, Update, VAC, Ongoing work to get the FRG gauges readouts to EPICs channels XGS600_calibration.pdf
Message ID: 16683     Entry time: Sat Feb 26 15:45:14 2022     In reply to: 16682     Reply to this: 16688   16691
Author: Tega 
Type: Update 
Category: VAC 
Subject: Ongoing work to get the FRG gauges readouts to EPICs channels 

I have attached a flow diagram of my understanding of how the gauges are connected to the network.

Earlier today, I connected the XGS-600 gauge controller to the IOLAN Serial Device Server on port 192.168.114.22 .

The plan is a follows:

1. Update the serial device yaml file to include this new ip entry for the XGS-600 gauge controller

2. Create a serial gauge class "serial_gauge_xgs.py" for the XGS-600 gauge controller that inherits from the serial gauge parent class for EPICS communication with a serial device via TCP sockets.

  • Might be better to use the current channels of the devices that are being replaced initially, i.e.
  • C1:Vac-FRG1_pressure C1:Vac-CC1_pressure
    C1:Vac-FRG2_pressure C1:Vac-CCMC_pressure
    C1:Vac-FRG3_pressure C1:Vac-PTP1_pressure
    C1:Vac-FRG4_pressure C1:Vac-CC4_pressure
    C1:Vac-FRG5_pressure C1:Vac-IG1_pressure

3. Modify the launcher file to include the XGS gauge controller. Following the same pattern used  to start the service for the other serial gauges, we can start the communication between the XGS-600 gauge controller and the IOLAN serial server and write data to EPICS channels using

controls@c1vac> python launcher.py XGS600

If we are able to establish communication between the XGS-600 gauge controller and write it gause data to EPICS channels, go on to steps 4.

4. Create a serial service file "serial_XGS600.service" and place it in the service folder

5. Add the new EPICS channels to the database file

6. Add the "serial_XGS600.service" to line 10 and 11 of modbusIOC.service

7. Later on, when we are ready, we can restart the updated modbusIOC service

 

For vacuum signal flow and Acromag channel assignments see [1]  and [2] respectively. For the 16 port IOLAN SDS (Serial Device Server) ethernet connections, see [3]. 

[1] https://wiki-40m.ligo.caltech.edu/Vacuum-Upgrade-2018?action=AttachFile&do=view&target=40m_Vacuum_System_Signal_Flow.pdf

[2] https://wiki-40m.ligo.caltech.edu/Vacuum-Upgrade-2018?action=AttachFile&do=view&target=AcromagChannelAssignment.pdf

[3] https://git.ligo.org/40m/vac/-/blob/master/python/serial/serial_devices.yaml

Attachment 1: Vac-gauges-flow-diagram.png  45 kB  Uploaded Sat Feb 26 17:31:00 2022  | Hide | Hide all
Vac-gauges-flow-diagram.png
ELOG V3.1.3-