The unit was tested before by Tege. The test included testing the testpoint voltages only. He summarized his work in this doc. The board number is S2100737. Here are the two comments about it:
"This unit presented with an issue on the PD1 circuit of channel 1-4 PCB where the voltage reading on TP6, TP7 and TP8 are -15.1V, -14.2V, and +14.7V respectively, instead of ~0V. The unit also has an issue on the PD2 circuit of channel 1-4 PCB because the voltage reading on TP7 and TP8 are -14.2V, and +14.25V respectively, instead of ~0V."
"Debugging showed that the opamp, AD822ARZ, for PD2 circuit was not working as expected so we replaced with a spare and this fixed the problem. Somehow, the PD1 circuit no longer presents any issues, so everything is now fine with the unit."
Note: No issues were reported on PD3 circuit is is malfunctioning now.
Quote: |
Also: Was this unit tested before? If so, what was the testing result at the time?
|
|