Message ID: 1651     Entry time: Thu Jun 4 15:53:15 2009
Author: steve 
Type: Update 
Category: PSL 
Subject: NPRO cooling flowrate adjusted 

The Neslab chiller is working well. It's temp display shows 20.0 C rock solid. Flow meter rotating at 13.5Hz at the out put of the chiller.

The MOPA temp was measured with a hand held thermocouple . The  PA was  34 C and 29 C at NPRO heat sink.

The NPRO flow meter was not rotating at this time. There was just trickeling water flow though the meter.

I closed the needle valve this point. It needed 8 turns clockwise. This drives head temp to 19.9 C

Than I opened the needle valve 9 turns and the flow meter wheel was  rotaing at ~ 1 Hz

We gained a little power. Can you explain this?

 

Attachment 1: needlevalve.jpg  86 kB  | Hide | Hide all