Entry  Wed Sep 29 17:10:09 2021, Anchal, Summary, CDS, c1teststand problems summary c1teststand_issues_summary.pdf
    Reply  Thu Sep 30 14:09:37 2021, Anchal, Summary, CDS, New way to ssh into c1teststand 
    Reply  Mon Oct 4 11:05:44 2021, Anchal, Summary, CDS, c1teststand problems summary 
       Reply  Mon Oct 4 18:00:16 2021, Koji, Summary, CDS, c1teststand problems summary 
          Reply  Tue Oct 5 17:58:52 2021, Anchal, Summary, CDS, c1teststand problems summary 
       Reply  Tue Oct 5 18:00:53 2021, Anchal, Summary, CDS, c1teststand time synchronization working now 
          Reply  Mon Oct 11 17:31:25 2021, Anchal, Summary, CDS, Fixed mounting of mx devices in fb. daqd_dc is running now. 
             Reply  Mon Oct 11 18:29:35 2021, Anchal, Summary, CDS, Moving forward? 
                Reply  Tue Oct 12 17:20:12 2021, Anchal, Summary, CDS, Connected c1sus2 to martian network 
                   Reply  Tue Oct 12 23:42:56 2021, Koji, Summary, CDS, Connected c1sus2 to martian network 
                      Reply  Wed Oct 13 11:25:14 2021, Anchal, Summary, CDS, Ran c1sus2 models in martian CDS. All good! CDS_screens_running.png
                         Reply  Tue Oct 19 18:20:33 2021, Ian MacMillan, Summary, CDS, c1sus2 DAC to ADC test data3_Plots.pdfdata2_Plots.pdf
                            Reply  Tue Oct 19 23:43:09 2021, Koji, Summary, CDS, c1sus2 DAC to ADC test P_20211019_224433.jpgP_20211019_224122.jpgP_20211019_224400.jpgP_20211019_224411.jpg
                               Reply  Wed Oct 20 11:48:27 2021, Anchal, Summary, CDS, Power supple configured correctly. 
                               Reply  Tue Oct 26 18:24:00 2021, Ian MacMillan, Summary, CDS, c1sus2 DAC to ADC test data2_Plots.pdfdata3_Plots.pdf
             Reply  Tue Oct 12 17:10:56 2021, Anchal, Summary, CDS, Some more information 
Message ID: 16398     Entry time: Wed Oct 13 11:25:14 2021     In reply to: 16397     Reply to this: 16414
Author: Anchal 
Type: Summary 
Category: CDS 
Subject: Ran c1sus2 models in martian CDS. All good! 

Three extra steps (when adding new models, new FE):

  • Chris pointed out that the sudo command in c1sus2 is giving error
    sudo: unable to resolve host c1sus2
    
    This error comes in when the computer could not figure out it's own hostname. Since FEs are network booted off the fb1, we need to update the /etc/hosts in /diskless/root everytime we add a new FE.
    controls@fb1:~ 0$ sudo chroot /diskless/root
    fb1:/ 0# sudo nano /etc/hosts
    fb1:/ 0# exit
    
    I added the following line in /etc/hosts file above:
    192.168.113.92  c1sus2 c1sus2.martian
    
    This resolved the issue of sudo giving error. Now, the rtcds make and install steps had no errors mentioned in their outputs.
  • Another thing that needs to be done, as Koji pointed out, is to add the host and models in /etc/rtsystab in /diskless/root of fb:
    controls@fb1:~ 0$ sudo chroot /diskless/root
    fb1:/ 0# sudo nano /etc/rtsystab
    fb1:/ 0# exit
    
    I added the following lines in /etc/rtsystab file above:
    c1sus2   c1x07  c1su2
    
    This told rtcds what models would be available on c1sus2. Now rtcds list is displaying the right models:
    controls@c1sus2:~ 0$ rtcds list
    c1x07
    c1su2
  • The above steps are still not sufficient for the daqd_ processes to know about the new models. This part is supossed to happen automatically, but does not happen in our CDS apparently. So everytime there is a new model, we need to edit the file /opt/rtcds/caltech/c1/chans/daq/master and add following lines to it:
    # Fast Data Channel lists
    # c1sus2
    /opt/rtcds/caltech/c1/chans/daq/C1X07.ini
    /opt/rtcds/caltech/c1/chans/daq/C1SU2.ini
    
    # test point lists
    # c1sus2
    /opt/rtcds/caltech/c1/target/gds/param/tpchn_c1x07.par
    /opt/rtcds/caltech/c1/target/gds/param/tpchn_c1su2.par
    
    I needed to restart the daqd_ processes in  fb1 for them to notice these changes:
    controls@fb1:~ 0$ sudo systemctl restart daqd_*
    
    This finally lit up the status channels of DC in C1X07_GDS_TP.adl and C1SU2_GDS_TP.adl . However the channels C1:DAQ-DC0_C1X07_STATUS and C1:DAQ-DC0_C1SU2_STATUS both have values 0x2bad. This persists on restarting the models. I then just simply restarted teh mx_stream on c1sus2 and boom, it worked! (see attached all green screen, never seen before!)

So now Ian can work on testing the I/O chassis and we would be good to move c1sus2 FE and I/O chassis to 1Y3 after that. I've also done following extra changes:

  • Updated CDS_FE_STATUS medm screen to show the new c1sus2 host.
  • Updated global diag rest script to act on c1xo7 and c1su2 as well.
  • Updated mxstream restart script to act on c1sus2 as well.
Attachment 1: CDS_screens_running.png  32 kB  Uploaded Wed Oct 13 12:59:13 2021  | Hide | Hide all
CDS_screens_running.png