Mon Sep 20 12:20:31 2021, Paco, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements   
|
Mon Sep 20 14:22:00 2021, rana, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements
|
Thu Sep 23 15:29:11 2021, Paco, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements
|
Fri Oct 1 14:25:27 2021, yehonathan, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements 
|
Mon Oct 4 16:00:57 2021, Yehonathan, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements  
|
Mon Oct 4 16:10:09 2021, rana, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements
|
Tue Oct 5 20:04:22 2021, Paco, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements
|
Wed Oct 6 15:39:29 2021, Anchal, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements
|
Tue Oct 12 11:32:54 2021, Yehonathan, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements 
|
Tue Oct 12 16:39:52 2021, rana, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements
|
Thu Oct 14 13:40:49 2021, Yehonathan, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements 
|
|
Message ID: 16374
Entry time: Mon Oct 4 16:00:57 2021
In reply to: 16358
Reply to this: 16375
16383
16393
|
Author: |
Yehonathan |
Type: |
Summary |
Category: |
SUS |
Subject: |
PRM and BS Angular Actuation transfer function magnitude measurements |
|
|
{Yehonathan, Anchel}
In an attempt to fix the actuation of the PRMI DOFs we set to modify the output matrix of the BS and PRM such that the response of the coils will be similar to each other as much as possible.
To do so, we used the responses at a single frequency from the previous measurement to infer the output matrix coefficients that will equilize the OpLev responses (arbitrarily making the LL coil as a reference). This corrected the imbalance in BS almost completely while it didn't really work for PRM (see attachment 1).
The new output matrices are shown in attachment 2-3. |
|
|
|