40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Sep 20 12:20:31 2021, Paco, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements PRM_PIT_ACT_TF.pdfPRM_YAW_ACT_TF.pdfBS_PIT_ACT_TF.pdfBS_YAW_ACT_TF.pdf
    Reply  Mon Sep 20 14:22:00 2021, rana, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements 
       Reply  Thu Sep 23 15:29:11 2021, Paco, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements BS_PRM_ANG_ACT_TF.pdf
          Reply  Fri Oct 1 14:25:27 2021, yehonathan, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements MICH_PRM_-0.34.pngMICH_PRM_-0.1.png
          Reply  Mon Oct 4 16:00:57 2021, Yehonathan, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements BS_PRM_ANG_ACT_TF_20211004.pdfBS_out_mat_20211004.txtPRM_out_mat_20211004.txt
             Reply  Mon Oct 4 16:10:09 2021, rana, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements 
             Reply  Tue Oct 5 20:04:22 2021, Paco, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements 
                Reply  Wed Oct 6 15:39:29 2021, Anchal, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements 
             Reply  Tue Oct 12 11:32:54 2021, Yehonathan, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements modified_output_matrices_radar_plots.pngnormal_output_matrices_radar_plots.png
                Reply  Tue Oct 12 16:39:52 2021, rana, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements 
                   Reply  Thu Oct 14 13:40:49 2021, Yehonathan, Summary, SUS, PRM and BS Angular Actuation transfer function magnitude measurements BSPRM_Actuation_Radar_plots.pngITM_Actuation_Radar_plots.png
Message ID: 16345     Entry time: Mon Sep 20 14:22:00 2021     In reply to: 16343     Reply to this: 16358
Author: rana 
Type: Summary 
Category: SUS 
Subject: PRM and BS Angular Actuation transfer function magnitude measurements 

I suggest plotting all the traces in the plot so we can see their differences. Also remove the 1/f^2 slope so that we can see small differences. Since the optlev servos all have low pass filters around 15-20 Hz, its not necessary to turn off the optlev servos for this measurement.

I think that based on the coherence and the number of averages, you should also be able to use Bendat and Piersol so estimate the uncertainy as a function of frequency. And we want to see the comparison coil-by-coil, not in the DoF basis.

4 sweeps for BS and 4 sweeps for PRM.