Tue May 26 17:05:44 2009, rob, Update, PSL, MOPA re-activated 
|
Tue May 26 17:34:14 2009, rob, Update, PSL, MOPA re-deactivated
|
Wed May 27 10:54:09 2009, rob, Update, PSL, we don't understand the chiller (broken)
|
Wed May 27 15:59:44 2009, rob, Update, PSL, we don't understand the chiller (broken)
|
Thu May 28 14:34:25 2009, rob, Update, PSL, chiller diagnosis
|
|
Message ID: 1629
Entry time: Thu May 28 14:34:25 2009
In reply to: 1628
|
Author: |
rob |
Type: |
Update |
Category: |
PSL |
Subject: |
chiller diagnosis |
|
|
Quote: |
steve, alberto, rob
After some futzing around with the chiller, we have come to the tentative conclusion that the refrigeration unit is not working. Steve called facilities to try to get them to recharge the refrigerant (R-404a) tomorrow, and we're also calling around for a spare chiller somewhere in the project (without luck so far).
|
The repair man thinks it's a bad start capacitor, which is 240uF at 120V. Steve has ordered a new one which should be here tomorrow, and with luck we'll have lasing by tomorrow afternoon. |