40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Mar 19 19:15:13 2021, Jon, Update, CDS, c1auxey assembly 
    Reply  Thu Mar 25 12:11:53 2021, Yehonathan, Update, CDS, c1auxey assembly 20210325_115500_HDR.jpg20210325_123033.jpg
       Reply  Thu Mar 25 14:16:33 2021, gautam, Update, CDS, c1auxey assembly 
       Reply  Tue Mar 30 17:27:04 2021, Yehonathan, Update, CDS, c1auxey assembly 
          Reply  Wed Aug 25 11:48:48 2021, Yehonathan, Update, CDS, c1auxey assembly 
             Reply  Mon Sep 13 14:32:25 2021, Yehonathan, Update, CDS, c1auxey assembly ETMX_OSEMS_Noise.png
                Reply  Wed Sep 15 11:27:50 2021, Yehonathan, Update, CDS, c1auxey assembly c1auxex_Grounding_OSEM_comparison1.pdfc1auxex_Grounding_OSEM_comparison2.pdf
       Reply  Thu Apr 15 17:24:08 2021, Jon, Update, CDS, Updated c1auxey wiring plan C1AUXEY_Chassis_Feedthroughs_-_By_Connector.pdf
          Reply  Mon Apr 19 21:54:55 2021, Yehonathan, Update, CDS, Updated c1auxey wiring plan 
             Reply  Wed Apr 28 18:56:57 2021, Yehonathan, Jon, Update, CDS, Updated c1auxey wiring plan 
                Reply  Thu Apr 29 16:35:51 2021, Yehonathan, Update, CDS, Updated c1auxey wiring plan 
                   Reply  Thu Apr 29 19:55:45 2021, Yehonathan, Update, CDS, Updated c1auxey wiring plan 
                      Reply  Fri Apr 30 19:18:51 2021, Yehonathan, Jon, Update, CDS, Updated c1auxey wiring plan 
                         Reply  Mon May 3 20:36:46 2021, Yehonathan, Jon, Update, CDS, Updated c1auxey wiring plan 
                            Reply  Sun May 23 18:28:54 2021, Jon, Update, CDS, Opto-isolator for c1auxey 
                               Reply  Fri May 28 10:54:59 2021, Jon, Update, CDS, Opto-isolator for c1auxey optoisolator.jpeg
                                  Reply  Thu Jun 3 17:15:17 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_Isolator_NPN_Input.pngOptical_Isolator_NPN_Output.png
                                     Reply  Thu Jun 3 22:08:00 2021, Koji, Update, CDS, Opto-isolator for c1auxey 
                                        Reply  Fri Jun 4 14:49:23 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_isolator_Wiring.pdf
                                           Reply  Fri Jun 4 17:46:25 2021, unYehonathan, Update, CDS, Opto-isolator for c1auxey 20210604_173420.jpg
                                              Reply  Sun Jun 6 16:33:47 2021, Jon, Update, CDS, BI channels on c1auxey 
                                                 Reply  Mon Jun 7 13:14:20 2021, Yehonathan, Update, CDS, BI channels on c1auxey 
                                                    Reply  Tue Jun 8 11:54:39 2021, Yehonathan, Update, CDS, BI channels on c1auxey 
                                           Reply  Sun Jun 6 03:02:14 2021, Koji, Update, CDS, Opto-isolator for c1auxey 
                                              Reply  Sun Jun 6 12:15:16 2021, Jon, Update, CDS, Opto-isolator for c1auxey 
                                              Reply  Sun Jun 6 15:59:51 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_isolator_Wiring.pdf
                                                 Reply  Mon Jun 14 15:31:30 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_isolator_Wiring.pdf
                                                    Reply  Tue Jun 15 21:48:55 2021, Koji, Update, CDS, Opto-isolator for c1auxey Acromag_RTS_BI_config.jpg
                                                       Reply  Wed Jun 16 17:24:29 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey Optical_isolator_Wiring.pdf
                                                          Reply  Wed Jun 16 20:32:39 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey 
                                                             Reply  Fri Jul 9 18:35:32 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey 
                                                                Reply  Mon Jul 12 18:06:25 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey 
                                                                   Reply  Wed Jul 28 12:47:52 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey SUS-ETMY_SparePDMon0_NoRef.pngSUS-ETMY_SparePDMon0_Ref_WithGND.pngDifferentialOutputTest.png
                                                                      Reply  Wed Aug 11 12:06:40 2021, Yehonathan, Update, CDS, Opto-isolator for c1auxey SUS-ETMY_SparePDMon0_2.png
Message ID: 16263     Entry time: Wed Jul 28 12:47:52 2021     In reply to: 16244     Reply to this: 16276
Author: Yehonathan 
Type: Update 
Category: CDS 
Subject: Opto-isolator for c1auxey 

To simulate a differential output I used two power supplies connected in series. The outer connectors were used as the outputs and the common connector was connected to the ground and used as a reference. I hooked these outputs to one of the differential analog channels and measured it over time using Striptool. The setup is shown in attachment 3.

I tested two cases: With reference disconnected (attachment 1), and connected (attachment 2). Clearly, the non-referred case is way too noisy.

Attachment 1: SUS-ETMY_SparePDMon0_NoRef.png  47 kB  Uploaded Wed Jul 28 13:54:27 2021  | Hide | Hide all
SUS-ETMY_SparePDMon0_NoRef.png
Attachment 2: SUS-ETMY_SparePDMon0_Ref_WithGND.png  38 kB  Uploaded Wed Jul 28 14:15:32 2021  | Hide | Hide all
SUS-ETMY_SparePDMon0_Ref_WithGND.png
Attachment 3: DifferentialOutputTest.png  38 kB  Uploaded Wed Jul 28 14:54:06 2021  | Hide | Hide all
DifferentialOutputTest.png
ELOG V3.1.3-