40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Thu Apr 29 11:51:27 2021, Anchal, Summary, LSC, Start of measuring IMC WFS noise contribution in ar cavity length noise 
    Reply  Thu Apr 29 15:11:33 2021, gautam, Update, CDS, RFM  RFM_errs.pngScreenshot_2021-04-29_15-12-56.png
       Reply  Thu Apr 29 17:43:16 2021, Koji, Update, CDS, RFM  
       Reply  Thu Apr 29 17:43:48 2021, Anchal, Update, CDS, F2A Filters double check F2AFiltersON.png
          Reply  Fri Apr 30 00:20:30 2021, gautam, Update, CDS, F2A Filters double check 
    Reply  Thu Apr 29 17:51:19 2021, Anchal, Summary, LSC, Start of measuring IMC WFS noise contribution in arm cavity length noise WFS1_PIT_exc_80-100Hz_Arms_ASD.pdf
       Reply  Fri Apr 30 00:18:40 2021, gautam, Summary, LSC, Start of measuring IMC WFS noise contribution in arm cavity length noise 
       Reply  Mon May 3 09:14:01 2021, Anchal, Paco, Update, LSC, IMC WFS noise contribution in arm cavity length noise WFS1_PIT_Noise_Inj_Test_IMC_unlocked.pdfWFS1_PIT_Noise_Inj_Test_ARM_locked.pdf
          Reply  Mon May 3 17:28:58 2021, Anchal, Paco, Rana, Update, LSC, IMC WFS noise contribution in arm cavity length noise excitationoftheMCanglessothatwecanseesomethingdotpng.pngexcitationoftheMCanglessothatwecanseesomethingdotpngbutthistimeitsMC2.png
             Reply  Tue May 4 11:43:09 2021, Anchal, Paco, Update, LSC, IMC WFS noise contribution in arm cavity length noise ArmCavNoiseContributions.pdfIOO-MC1_PIT_NoiseInjTest2.pdfIOO-MC1_PIT_NoiseInjTest_AWGGUI_Config.png
                Reply  Fri May 7 11:54:02 2021, Anchal, Paco, Update, LSC, IMC WFS noise contribution in arm cavity length noise ITMX-XARM_TF.pdfITMY-YARM_TF.pdfArmCavNoiseContributions.pdf
                   Reply  Mon May 10 18:19:12 2021, Anchal, Paco, Update, LSC, IMC WFS noise contribution in arm cavity length noise, Corrections ArmCavNoiseContributions.pdf
                      Reply  Wed May 12 10:53:20 2021, Anchal, Paco, Update, LSC, PSL-IMC PDH Loop and XARM PDH Loop diagram IMC_SingleArm.pdf
Message ID: 16101     Entry time: Thu Apr 29 17:51:19 2021     In reply to: 16095     Reply to this: 16104   16108
Author: Anchal 
Type: Summary 
Category: LSC 
Subject: Start of measuring IMC WFS noise contribution in arm cavity length noise 

t Both arms were locked simply by using IFO > Configure > ! (YARM) > Restore YARM. I had to use ASS to improve the TRX/TRY to ~0.95.

I measured C1:LSC-XARM_IN1_DQ and C1:LSC-YARM_IN1_DQ while injecting band limited noise in C1:IOO-WFS1_PIT_EXC using uniform noise with amplitude 1000 along with filter defined by string: cheby1("BandPass",4,1,80,100). I calibrated the control arms signals by 2.44 nm/cts calibration factor directly picked up from 13984.

For the duration of this test, all LIMIT switches in the WFS loops were switched OFF.

I do not see any affect on the arm control signal power spectrums with or without the noise injection. Attachement 1 shows the PSD along with PSD of the injection site IN2 signal. I must be doing something wrong, so would like feedback before I go further.

Attachment 1: WFS1_PIT_exc_80-100Hz_Arms_ASD.pdf  10 kB  Uploaded Thu Apr 29 18:56:45 2021  | Hide | Hide all
WFS1_PIT_exc_80-100Hz_Arms_ASD.pdf
ELOG V3.1.3-