Message ID: 16085     Entry time: Mon Apr 26 18:52:52 2021
Author: Anchal, Paco 
Type: HowTo 
Category: Computer Scripts / Programs 
Subject: awg free slot 

Today we had some trouble launching an excitation on C1:IOO-MC_LSC_EXC from awggui. The error read:

awgSetChannel: failed getIndexAWG C1:SUS-MC2_LSC_EXC ret=-3

What solved this was the following :

  1. launch the dtt command line interface
  2. Anchal remembers a slot number 37008
  3. We issue >> awg free 37008
  4. Slot freed, launch a new instance of awggui
ELOG V3.1.3-