Message ID: 16046     Entry time: Sun Apr 18 21:29:55 2021     In reply to: 16044     Reply to this: 16062
Author: rana 
Type: Update 
Category: PSL 
Subject: Laser amplifier 
  • Ideally, we put the chiller outside of the interferometer area. The PSL chiller used to be in the control room near the door by IMC REFL. We could also put it in the drill press room.
  • Once we figure out a couple of places where the Diode Box can go, we can ask facilities to make the appropriate power connections. They will have to eval the situation to figure out if the main power to the lab needs to be shut down.
  • Can we put the laser diode box in the drill press room too? Then the hoses can be short. Perhaps less EMI getting into our sensitive places.