I finished testing the OSEMs. I put all the OSEMs back in the box. The OSEMS were divided into several bags. I put the OSEM box next to the south flow bench on the floor.
I have uploaded the OSEM catalog to the wiki. I will upload the LED spot images later.
In summary:
Total 64 OSEMS, 31 long, 33 short.
Perfectly centered LED spots, ready for C&B OSEMS: 30, 12 long, 18 short.
Perfectly centered LED spots, need some work (missing pigtails, weird screws) OSEMS: 7, 5 long, 2 short.
Slightly off-centered (subjective) LED spots, ready for C&B OSEMS: 20, 7 long, 13 short.
Slightly off-centered (subjective) LED spots, need some work (missing pigtails, weird screws) OSEMS: 4 long
Defective OSEMS or LED spot way off-center: 3. |