On Friday evening we checked out a few more things, somewhat overlapping with previous tests. All tests done with PRMI on carrier lock (REFL11_I -> PRC, AS55_Q-> MICH):
- check that PRC drive appropriately minimizes in REFL55_Q. I:Q ratio is ~100:1; good enough.
- put sine waves around 311 and 333 Hz into PRCL and MICH at the LSC output matrix using awggui and LSC osc. not able to adjust LSC/OSC output matrix to minimize the MICH drive in REFL_I.
- measured the TF from BS & PRM LSC drive to the REFL55_I/Q outputs. very nearly the same audio frequency phase, so the problem is NOT in the electronics || mechanical transfer functions of the suspensions.
Further questions:
- is this something pathological in the PRMI carrier lock? we should check by locking on sidebands to REFL55 and REFL165 and repeat tests.
- Can it be a severe mode mismatch from IMC output to PRMI mode? the cavity should be stable with the flipped folding mirrors, but maybe something strange happening. How do we measure the mode-matching to the PRC quantitatively?
- huge RAM is ruled out by Gautam's test of looking at REFL demod signals: dark offset vs. offset with a single bounce off of PRM (with ITMs mis-aligned)
- if there is a large (optical) offset in the AS55_Q lock point, how big would it have to be to mess up the REFL phase so much?
- what is going on with the REFL55 whitening/AA electronics?
unrelated note: Donatella the Workstation was ~3 minutes ahead of the FE machines (you can look at the C0:TIM-PACIFIC_STRING on many of the MEDM screens for a rough simulacrum). When the workstation time is so far off, DTT doesn't work right (has errors like test timed out, or other blah blah). I installed NTP on donatella and started the service per SL7 rules. Since we want to migrate all the workstations to Debian (following the party line), lets not futz with this too much.
gautam, 1 Mar 1600: In case I'm being dumb, I attach the screen grab comparing dark offset to the single bounce off PRM, to estimate the RAM contribution. The other signals are there just to show that the ITMs are sufficiently misaligned. The PRCL PDH fringe is usually ~12000 cts in REFL11, ~5000cts in REFL55, and so the RAM offset is <0.1% of the horn-to-horn PDH fringe.
P.S. I know generally PNGs in the elog are frowned upon. But with so many points, the vector PDF export by NDS (i) is several megabytes in size and (ii) excruciatingly slow. I'm proposing a decimation filter for the export function of ndscope - but until then, I claim plotting with "rasterized=True" and saving to PDF and exporting to PNG are equivalent, since both yield a rasterized graphic. |