Entry  Wed Jan 20 18:07:09 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements before modifications 7x
    Reply  Mon Jan 25 18:18:04 2021, Anchal, Summary, BHD, Satellite Amplifier Transfer Functions and noise D1002818_S2100029_TFandNoiseSpectrum.pdfD1002818_Testing.zip
       Reply  Wed Feb 3 16:27:19 2021, Anchal, Summary, BHD, Satellite Amplifier Transfer Functions and noise After modifications D1002818_S2100029_TFAfterChanges.pdfD1002818_S2100029_OutputNoiseSpecAfterChanges.pdfD1002818_S2100029_InputRefferedNoiseSpecAfterChanges.pdfD1002812_S2100029_After_Modifications_Feb3.jpgAfterChanges.zip
          Reply  Wed Feb 10 11:45:59 2021, Anchal, Summary, BHD, Satellite Amplifier Very Low frequency noise After modifications 6x
             Reply  Wed Feb 10 15:07:50 2021, Anchal, Summary, BHD, Satellite Amplifier Output Offset measurements 
                Reply  Wed Feb 10 15:25:45 2021, gautam, Summary, BHD, Satellite Amplifier Output Offset measurements 
                   Reply  Wed Feb 10 17:18:03 2021, Koji, Summary, BHD, Satellite Amplifier Output Offset measurements 
             Reply  Thu Feb 11 11:10:05 2021, Anchal, Summary, BHD, Satellite Amplifier Very Low frequency noise After modifications AfterChangesSpectrum_AC.zipD1002818_S2100029_OutputNoiseSpecAfterChanges.pdfD1002818_S2100029_InputRefCurrentNoiseSpecAfterChanges.pdfD1002818_S2100029_InputRefDispNoiseSpecAfterChanges.pdf
                Reply  Thu Feb 11 16:58:52 2021, rana, Summary, BHD, Satellite Amplifier Very Low frequency noise After modifications 
    Reply  Thu Jan 28 12:53:14 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements before modifications D1100117_S2100027_Current_Noise_Spectrum.pdfD1100117_S2100027_Voltage_Noise_Spectrum.pdfD1100117_S2100028_Current_Noise_Spectrum.pdfD1100117_S2100028_Voltage_Noise_Spectrum.pdfSpectrumMeasurement.zip
       Reply  Thu Jan 28 18:04:55 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements After modifications D1100117_S2100027_TF.pdfAfterChanges.zip
          Reply  Thu Jan 28 21:44:45 2021, gautam, Summary, BHD, HAM-A Coil Driver measurements After modifications 
             Reply  Fri Jan 29 15:39:30 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements After modifications TF and Noise S2100027 D1100687_S2100027_After_Modifications_Jan28.jpgD1100117_S2100027_TF.pdfD1100117_S2100027_Voltage_Noise_Spectrum.pdfD1100117_S2100027_Current_Noise_Spectrum.pdfAfterChanges.zip
                Reply  Tue Feb 2 11:57:46 2021, Anchal, Summary, BHD, HAM-A Coil Driver measurements After modifications TF and Noise S2100028 D1100687_S2100028_After_Modifications_Feb01_2021.jpgD1100117_S2100028_TF.pdfD1100117_S2100028_Voltage_Noise_Spectrum.pdfD1100117_S2100028_Current_Noise_Spectrum.pdfAfterChanges.zip
Message ID: 15797     Entry time: Wed Feb 10 11:45:59 2021     In reply to: 15793     Reply to this: 15799   15803
Author: Anchal 
Type: Summary 
Category: BHD 
Subject: Satellite Amplifier Very Low frequency noise After modifications 

As suggested, I wrapped the satellite amplifier box D10028128 S2100029 in blanket and foam and took very low frequency spectrum starting from 32 mHz to 3 Hz. The results are attached along with stiched high frequency measurements from 40m/15793.

Very Low Frequency Spectrum Measurement

  • D1002818 S2100029 box was powered and covered in a foam blanket.
  • Additionally, it was covered from all sides with foam to reduce wind and temperature effects on it.
  • The rear panel DB25 connector was connected to a breakout board where pins od PDA input and GND were shorted, shorting the transimpedance circuit input.
  • The output was read from PDMon DB9 output at front panel which was converted to 4 BNC channels using breakout board.
  • Two channel noise was measured at once using D1002818_SP.yml parameter file.
  • Instrument noise at all the used input ranges were measured separately by shorting the input of the BNC cables.

Edit Wed Feb 10 15:14:13 2021 :

THIS MEASUREMENT WAS WRONG. SEE 40m/15799.

Attachment 1: FrontsideLook.jpg  854 kB  | Hide | Hide all
FrontsideLook.jpg
Attachment 2: BacksideLook.jpg  895 kB  | Hide | Hide all
BacksideLook.jpg
Attachment 3: InnerFoamBlanket.jpg  1.103 MB  | Hide | Hide all
InnerFoamBlanket.jpg
Attachment 4: D1002818_S2100029_OutputNoiseLFSpecAfterChanges.pdf  244 kB  | Hide | Hide all
D1002818_S2100029_OutputNoiseLFSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseLFSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseLFSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseLFSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseLFSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseLFSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseLFSpecAfterChanges.pdf D1002818_S2100029_OutputNoiseLFSpecAfterChanges.pdf
Attachment 5: D1002818_S2100029_InputRefCurrentNoiseLFSpecAfterChanges.pdf  239 kB  | Hide | Hide all
D1002818_S2100029_InputRefCurrentNoiseLFSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseLFSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseLFSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseLFSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseLFSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseLFSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseLFSpecAfterChanges.pdf D1002818_S2100029_InputRefCurrentNoiseLFSpecAfterChanges.pdf
Attachment 6: AfterChangesLFSpectrum.zip  12.370 MB
ELOG V3.1.3-