40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Nov 18 22:49:06 2020, gautam, Update, ASC, Some thoughts about AS WFS electronics 
    Reply  Sun Nov 22 21:09:37 2020, gautam, Update, ASC, Planned mods for WFS head aLIGO_wfs_v5_40m.pdfTFs.pdfnoise.pdf
       Reply  Wed Nov 25 18:18:41 2020, gautam, Update, ASC, Planned mods for WFS head anodeVsOutput.pngsensitivity.png
          Reply  Sat Nov 28 21:44:53 2020, rana, Update, ASC, Planned mods for WFS head 
          Reply  Thu Dec 10 22:46:57 2020, gautam, Update, ASC, WFS head assembled aLIGO_wfs_v5_40m.pdfTF_meas.pdf
             Reply  Thu Dec 17 15:23:56 2020, gautam, Update, ASC, WFS head characterization oNoise.pdfZ_Hi.pdfZ_Low.pdfIMG_9030.jpgiNoise.pdf
    Reply  Wed Nov 25 18:30:23 2020, gautam, Update, ASC, Some thoughts about AS WFS electronics 
       Reply  Wed Dec 2 23:07:19 2020, gautam, Update, ASC, Electrical LO signal for AS WFS RF_Frequency_Source.pdf
          Reply  Thu Dec 3 11:02:35 2020, rana, Update, Electronics, Electrical LO signal for AS WFS 
          Reply  Thu Dec 3 13:55:52 2020, Koji, Update, ASC, Electrical LO signal for AS WFS 
             Reply  Thu Dec 3 20:38:46 2020, gautam, Update, ASC, Electrical LO signal for AS WFS RF_Frequency_Source.pdfdemodPath.pdfEOMpath.pdf
                Reply  Thu Dec 3 21:14:38 2020, Koji, Update, ASC, Electrical LO signal for AS WFS 
                   Reply  Thu Dec 3 21:44:49 2020, gautam, Update, ASC, Electrical LO signal for AS WFS 
                      Reply  Fri Dec 4 03:02:40 2020, Koji, Update, ASC, Electrical LO signal for AS WFS 
                         Reply  Fri Dec 4 22:41:56 2020, gautam, Update, ASC, Freq Gen Box revamp 
                            Reply  Sat Dec 5 20:44:35 2020, gautam, Update, ASC, Freq Gen Box re-installed 6x
                               Reply  Mon Dec 7 12:38:51 2020, gautam, Update, IOO, IMC loop char OLTFcomparison.pdfdemodMons.pdfOLTFcomparison.pdf
                               Reply  Mon Dec 7 22:54:30 2020, gautam, Update, LSC, Modulation depth measurement modDepth.pdf
                                  Reply  Sat Jan 16 18:59:44 2021, gautam, Update, LSC, Modulation depth measurement PLL.pdfmodDepth.pdfmodProfile.pdfEOMpath_postMod.pdf
Message ID: 15690     Entry time: Wed Nov 25 18:30:23 2020     In reply to: 15682     Reply to this: 15697
Author: gautam 
Type: Update 
Category: ASC 
Subject: Some thoughts about AS WFS electronics 

An 8 channel whitening chassis was prepared and tested. I measured:

  1. TF from input to output - there are 7 switchable stages (3 dB, 6 dB, 12 dB and 24 dB flat whitening gain, and 3 stages of 15:150 Hz z:p whitening). I enabled one at a time and measured the TF. 
  2. Noise with input terminated.

In summary,

  1. All the TFs look good (I will post the plots later), except that the 3rd stage of whitening on both boards don't show the expected transfer function. The fact that it's there on both boards makes me suspect that the switching isn't happening correctly (I'm using a little breakout board). I'm inclined to not debug this because it's unlikely we will ever use 3 stages of 15:150 whitening for the AS WFS. 
  2. The noise measurement displayed huge (x1000 above the surrounding broadband noise floor) 60 Hz harmonics out to several kHz. My hypothesis is that this has to do with some bad grounding. I found that the circuit ground is shorted to the chassis via the shell of the 9pin and 15pin Dsub connectors (but the two D37 connector shields are isolated). This seems very wierd, idk what to make of this. Is this expected? Looking at the schematic, it would appear that the shields of the connectors are shorted to ground which seems like a bad idea. afaik, we are using the same connectors as on the chassis at the sites - is this a problem there too? Any thoughts
Quote:

Whitening chassis. Waiting for front panels to arrive, PCBs and interface board are in hand, stuffed and ready to go. A question here is how we want to control the whitening - it's going to be rather difficult to have fast switchable whitening. I think we can just fix the whitening state. Another option would be to control the whitening using Acromag BIO channels.