Entry  Wed Sep 23 08:57:18 2020, gautam, Update, VAC, TP2 running HOT Screen_Shot_2020-09-23_at_8.55.39_AM.png
    Reply  Wed Sep 23 10:06:52 2020, Koji, Update, VAC, TP2 running HOT Screen_Shot_2020-09-23_at_10.00.43.png
Message ID: 15599     Entry time: Wed Sep 23 08:57:18 2020     Reply to this: 15600
Author: gautam 
Type: Update 
Category: VAC 
Subject: TP2 running HOT 

The interlocks tripped at ~630am local time. Jordan reported that TP2 was supposedly running at 52 C (!).

V1 was already closed, but TP2 was still running. With him standing by the rack, I remotely exectued the following sequence:

  • VM1 closed (isolates RGA volume).
  • VA6 closed (isolates annuli from being pumped).
  • V7 opened (TP3 now backs TP1, temporarily, until I'm in the lab to check things out further).
  • TP2 turned off.

Jordan confirmed (by hand) that TP2 was indeed hot and this is not just some serial readback issue. I'll do the forensics later.