40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Sat Sep 19 23:59:22 2020, anchal, Summary, ALS, ALS noise budget update ALS_nb_ExtraPoles.pdfALS_controls.yamlALS_nb_Kiwamus_Values.pdfALS_controls_Kiwamus_Values.yamlALS_simulink_model.svg
    Reply  Sun Sep 20 23:12:13 2020, rana, Summary, ALS, ALS noise budget update 
       Reply  Tue Sep 22 00:14:43 2020, anchal, Summary, ALS, ALS noise budget update ALS_NoiseBudgetUpdate.pdfALS_controls.yaml
          Reply  Tue Sep 22 12:14:42 2020, rana, Summary, ALS, ALS noise budget update 
             Reply  Wed Sep 23 11:13:49 2020, anchal, Summary, ALS, ALS noise budget update ALS_NoiseBudgetUpdate.pdfALS_controls.yaml
                Reply  Wed Oct 7 16:56:23 2020, anchal, Summary, ALS, ALS noise budget update - Updated AUX PDH Loop values Extract_X_AUX_PDH_Model.zipALS_NoiseBudgetUpdate.pdf
                   Reply  Thu Oct 8 11:59:52 2020, rana, Summary, ALS, ALS noise budget update - Updated AUX PDH Loop values 
                      Reply  Fri Oct 9 18:32:14 2020, anchal, Summary, ALS, ALS noise budget update - Updated AUX PDH Loop values 
                         Reply  Wed Oct 14 17:03:55 2020, anchal, Summary, ALS, ALS noise budget update - Added whitening filter for ADC ALS_NoiseBudgetUpdate.pdf
Message ID: 15594     Entry time: Tue Sep 22 12:14:42 2020     In reply to: 15593     Reply to this: 15601
Author: rana 
Type: Summary 
Category: ALS 
Subject: ALS noise budget update 

This ALS loop is not stable. Its one of those traps that comes from using only the Bode plot to estimate the loop stability. You have to also look at the time domain response - you can look at my feedback lecture for the SURF students for some functions.

ELOG V3.1.3-