See Attachment #1.
- The original ETMY actuation matrix was the naive one so I simply retuned everything by adding the butterfly mode appropriately.
- The cross coupling between the DOFs has not been characterized, but the local damping, Oplev loops, POX/POY locking, simpel Michelson locking, and ASS dither alignemnt all seem to work so I'm thinking this is good enough for now.
- A copy of the ETMYaux.db file was made, and the slow bias voltage was redistributed among the three available face OSEMs - note that the magnet polarity has to be taken into account.
- The series resistance on the slow path was reduced from 430 ohms, 1W to 100 ohms, 3W, to allow DC alignment of the cavity axis to the beam axis.
- Vertex Oplevs were re-centered on their respective QPDs after locking POX/POY and running the ASS dither alignment.
- The AS spot was a little low on the camera - presumably the SR2/SR3 got macroscopically misaligned. I re-centered the beam on the camera on the AS table (by moving the camera, the beam path was not disturbed).
The beam spot on ETMY looks weird (looks almost like a TEM10 mode), but the one on ITMY seems fine, see Attachment #2. Wonder what's going on there, maybe just a focusing problem?
Quote: |