Entry  Sat Sep 19 18:07:35 2020, Koji, Update, VAC, TP3 RP failure vac.pngScreen_Shot_2020-09-19_at_17.52.40.png
    Reply  Sat Sep 19 19:37:16 2020, not Koji, Update, VAC, TP3 RP failure 
Message ID: 15586     Entry time: Sat Sep 19 19:37:16 2020     In reply to: 15582
Author: not Koji 
Type: Update 
Category: VAC 
Subject: TP3 RP failure 

Disconcerting because those tip seals were just replaced [15417]. Maybe they were just defective, but if there is a more serious problem with the pump, there is a spare Varian roughing pump (the old TP2 dry pump) sitting at the X-end.

I reset the interlock error to unfreeze the vac controls (leaving V5 closed).

Quote:

So the conclusion is that RP for TP3 has failed. Presumably, the tip-seal needs to be replaced.

Right now TP3 was turned off and is ready for the tip-seal replacement. V5 was closed since the watchdog tripped.