Message ID: 15581     Entry time: Sat Sep 19 11:27:04 2020     In reply to: 15580     Reply to this: 15583
Author: rana 
Type: Update 
Category: General 
Subject: M4.5 EQ in LA 

the seismometers obviously saturated during the EQ, but the accelerometers captured some of it. It looks like there's different saturation levels on different sensors.

Also, it seems the mounting of the MC2 accelerometers is not so good. There's some ~10-20 Hz resonance its mount that's showing up. Either its the MC2 chamber legs or the accelerometers are clamped poorly to the MC2 baseplate.

Sun Sep 20 00:02:36 2020 edit: fixed indexing error in plots

* also assuming that the sensors are correctly calibrated in the front end to 1 count = 1 um/s^2 (this is what's used in the summ pages)

Attachment 1: Sep18-EQ.pdf  269 kB  Uploaded Sun Sep 20 01:03:01 2020  | Hide | Hide all
Sep18-EQ.pdf
ELOG V3.1.3-