Message ID: 15547     Entry time: Sat Aug 29 20:07:48 2020     In reply to: 15450     Reply to this: 15548   15572
Author: rana 
Type: Update 
Category: Electronics 
Subject: WFS characterization 

I set up to do the WFS head modifications today, but I was shot down in flames due to a missing AC/DC adapter.

The Prologix GPIB-ethernet dongle needs +8-13 V to run. Some riff raff has removed the adapter and I was thunderstruck to see that it had not been returned.

I did the usual hunt around the lab looking for something with the right specs and connector. I found one that could do +9V and had the right connector, but it didn't light up the adapter so I put it back in black SP table.

I'll order a couple of these (5 ordered for delivery on Wednesday) in case there's a hot demand for the jack / plug combo that this one has. The setup is in the walkway, but I returned the AS table to the usual state and made sure the IMC is locking well.