Entry  Thu Jun 11 11:48:43 2020, gautam, Update, VAC, Vac failure vacFailure.png
    Reply  Thu Jun 11 16:14:03 2020, gautam, Update, VAC, Vac failure - probable cause is serial comm glitch TP2_tempGlitch.pngPSL_shutterClosed.pngTP2tempGlitches.pdf
       Reply  Thu Jun 11 17:35:34 2020, gautam, Update, VAC, Pumpspool UPS needs battery replacement 
          Reply  Fri Jun 12 01:23:32 2020, Koji, Update, VAC, Pumpspool UPS needs battery replacement 
             Reply  Fri Jun 12 17:32:40 2020, Koji, Update, VAC, Pumpspool UPS needs battery replacement Screen_Shot_2020-06-12_at_17.22.23.png
                Reply  Fri Jun 12 19:02:52 2020, gautam, Update, VAC, Pumpspool UPS needs battery replacement 
                   Reply  Fri Jun 12 19:23:56 2020, Koji, Update, VAC, Pumpspool UPS needs battery replacement 
                      Reply  Fri Jun 12 19:33:31 2020, gautam, Update, VAC, Pumpspool UPS needs battery replacement 
                         Reply  Tue Jun 16 13:35:03 2020, Jon, Update, VAC, Temporary vac fix / IFO usable again Pumpdown.png
                            Reply  Wed Jun 17 16:27:51 2020, gautam, Update, VAC, Questions/comments on vacuum 
                               Reply  Thu Jun 18 11:00:24 2020, Jon, Update, VAC, Questions/comments on vacuum 
                                  Reply  Thu Jun 18 12:00:36 2020, gautam, Update, VAC, Questions/comments on vacuum 
                                     Reply  Thu Jun 18 14:13:03 2020, Jon, Update, VAC, Questions/comments on vacuum 
                                        Reply  Thu Jun 18 15:46:34 2020, gautam, Update, VAC, Questions/comments on vacuum 
                                           Reply  Fri Jun 19 07:40:49 2020, Jon, Update, VAC, Questions/comments on vacuum 
                                              Reply  Fri Jun 19 09:57:35 2020, gautam, Update, VAC, Questions/comments on vacuum 
Message ID: 15402     Entry time: Tue Jun 16 13:35:03 2020     In reply to: 15399     Reply to this: 15404
Author: Jon 
Type: Update 
Category: VAC 
Subject: Temporary vac fix / IFO usable again 

[Jon, Jordan, Koji]

Today Jordan reconfigured the vac system to allow pumping of the main volume resume, with Jon and Koji remotely advising. All clear to resume normal IFO activities. However, the vac system is operating in a temporary configuration that will have to be reverted as we locate replacement components. Details below.

Procedure

Since serial readback of the TP2 controller seems to be failing, we reconfigured the system with TP3 now backing for TP1. TP2 was valved off (at V4) and shut down until we can replace its controller.

TP3 has its own problems, however. It was valved off in January after its temperature readback began glitching and spuriously triggering the interlocks [ELOG 15140]. However the problem appears to be limited only one readback (rotation speed, current, voltage are fine) and there is enough redundancy in the pump-dependent interlock conditions to safely connect it to the main volume.

We also discovered that sometime since January, the TP3 dry pump has failed. The foreline pressure had risen to 165 torr. Since the TP2 and TP3 dry pumps are not interchangeable (Agilent vs. Varian), we instead valved in the auxiliary dry pump and disconnected the failed dry pump using a KF blank. This is a temporary arrangement until the permanent dry pump can be repaired. Jordan removed it to replace the tip seals and will test it in the bake lab before reinstalling.

With this configuration in place, we proceeded to pump down the main volume without issue (attachment 1). We monitored the pumpdown for about 45 min., until the pressure had reached ~1E-5 torr and TP3 had been transitioned to standby (low-speed) mode.

Summary of topology changes:

  • TP2 valved off and shut down until controller can be replaced
  • TP3 temporarily backing for TP1
  • Auxiliary dry pump temporarily backing for TP3
  • TP3 dry pump has been removed for repairs
Attachment 1: Pumpdown.png  31 kB  Uploaded Tue Jun 16 15:15:15 2020  | Hide | Hide all
Pumpdown.png
ELOG V3.1.3-