Entry  Thu Jun 11 11:48:43 2020, gautam, Update, VAC, Vac failure vacFailure.png
    Reply  Thu Jun 11 16:14:03 2020, gautam, Update, VAC, Vac failure - probable cause is serial comm glitch TP2_tempGlitch.pngPSL_shutterClosed.pngTP2tempGlitches.pdf
       Reply  Thu Jun 11 17:35:34 2020, gautam, Update, VAC, Pumpspool UPS needs battery replacement 
          Reply  Fri Jun 12 01:23:32 2020, Koji, Update, VAC, Pumpspool UPS needs battery replacement 
             Reply  Fri Jun 12 17:32:40 2020, Koji, Update, VAC, Pumpspool UPS needs battery replacement Screen_Shot_2020-06-12_at_17.22.23.png
                Reply  Fri Jun 12 19:02:52 2020, gautam, Update, VAC, Pumpspool UPS needs battery replacement 
                   Reply  Fri Jun 12 19:23:56 2020, Koji, Update, VAC, Pumpspool UPS needs battery replacement 
                      Reply  Fri Jun 12 19:33:31 2020, gautam, Update, VAC, Pumpspool UPS needs battery replacement 
                         Reply  Tue Jun 16 13:35:03 2020, Jon, Update, VAC, Temporary vac fix / IFO usable again Pumpdown.png
                            Reply  Wed Jun 17 16:27:51 2020, gautam, Update, VAC, Questions/comments on vacuum 
                               Reply  Thu Jun 18 11:00:24 2020, Jon, Update, VAC, Questions/comments on vacuum 
                                  Reply  Thu Jun 18 12:00:36 2020, gautam, Update, VAC, Questions/comments on vacuum 
                                     Reply  Thu Jun 18 14:13:03 2020, Jon, Update, VAC, Questions/comments on vacuum 
                                        Reply  Thu Jun 18 15:46:34 2020, gautam, Update, VAC, Questions/comments on vacuum 
                                           Reply  Fri Jun 19 07:40:49 2020, Jon, Update, VAC, Questions/comments on vacuum 
                                              Reply  Fri Jun 19 09:57:35 2020, gautam, Update, VAC, Questions/comments on vacuum 
Message ID: 15393     Entry time: Thu Jun 11 17:35:34 2020     In reply to: 15392     Reply to this: 15394
Author: gautam 
Type: Update 
Category: VAC 
Subject: Pumpspool UPS needs battery replacement 

Summary:

The pumpspool UPS has its "Replace Battery" indicator light on. Might be a good chance to change the UPS, but at the very least, we should put in fresh batteries (last replacement was in Aug 2017).

I'll say this again - the pumpspool area is noisier than I remember it being, I think one / both of the roughing pumps backing TP2 / TP3 need tip-seal replacements.

BTW, EX is 5C hotter than EY, by virtue of the tarnac outside? In fact, judging by Steve's thermometers, EX reports a 12C swing in 24 hours between 30 C and 18 C (so almost no temperature control) while EY reports a 5C swing between 20 and 25 C. This is borne out by the ETM Oplev data I think...