Message ID: 15329     Entry time: Wed May 13 15:13:11 2020     In reply to: 15323     Reply to this: 15332
Author: Yehonathan 
Type: Update 
Category: Loss Measurement 
Subject: 40m Phase maps loss estimation 

Koji pointed out during the group meeting that I should compensate for local tilt when I move the beam around the mirror for calculating the loss map.

So I did.

Also, I made a mistake earlier by calculating the loss map for a much bigger (X7) area than what I thought.

Both these mistakes made it seem like the loss is very inhomogeneous across the mirror.

Attachment 1 and 2 show the corrected loss maps for ITMX and ETMX respectively.

The loss now seems much more reasonable and homogeneous and the average total arm loss sums up to ~ 22ppm which is consistent with the after-cleaning arm loss measurements.