40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 15300     Entry time: Tue Apr 7 15:30:40 2020
Author: Jon 
Type: Summary 
Category: NoiseBudget 
Subject: 40m noise budget migrated to pygwinc 

In the past year, pygwinc has expanded to support not just fundamental noise calculations (e.g., quantum, thermal) but also any number of user-defined noises. These custom noise definitions can do anything, from evaluating an empirical model (e.g., electronics, suspension) to loading real noise measurements (e.g., laser AM/PM noise). Here is an example of the framework applied to H1.

Starting with the BHD review-era noises, I have set up the 40m pygwinc fork with a working noise budget which we can easily expand. Specific actions:

  • Updated the 40m fork to the latest pygwinc version (while preserving the commit history).
  • Added a directory ./CIT40m containing the 40m-specific noise budget files (created by GV).
  • Added an ipython notebook CIT40m.ipynb at the root level showing how to generate a noise budget.
  • Integrated our DAC and seismic noise estimators into pygwinc.
  • Marked the old 40m NB repo as obsolete (last commit > 2 yrs ago). Many of these noise estimates are probably stale, but I will work with GV to identify which ones can be migrated.

I set up our fork in this way to keep the 40m separate from the main pygwinc code (i.e., not added to as a built-in IFO type). With the 40m code all contained within one root-level directory (with a 40m-specific name), we should now always be able to upgrade to the latest pygwinc without creating intractable merge conflicts.