{Yehonathan, Jon, Jordan}
I tested the ai channels of the new PSL Acromag by looping an already-tested ao channel (C2:PSL-FSS-INOFFSET) back to the different ai channels.
I use Jon's IFOTest with /users/jon/ifotest/PSL.yaml.
I created a spreadsheet for the testing based on the current wiring spreadsheet. I added two columns for the high and low readings for each ai channel (attached pdf).
I marked in red the failed channels. Some of them are probably calibration issues, but the ones that show the same voltage for high and low are probably disconnected wires.
I redid the test on the channel that seemed disconnected to confirm.
I created a yaml file with all the failed channels for retesting called /users/jon/ifotest/PSL_failed_channels.yaml. |