Mon Jan 20 20:20:36 2020, gautam, Update, PSL, PMC servo checkout 
|
Wed Jan 22 11:22:39 2020, gautam, Update, PSL, PMC modulation depth measurement
|
Wed Jan 22 20:12:36 2020, gautam, Update, PSL, PMC demodulator electrical characterization 
|
Thu Jan 23 14:37:05 2020, gautam, Update, PSL, PMC VGA chip damaged?
|
Thu Jan 23 16:37:14 2020, rana, Update, PSL, PMC VGA chip damaged?
|
Thu Jan 23 22:10:01 2020, gautam, Update, PSL, PMC servo pulled out
|
Fri Jan 24 15:42:08 2020, gautam, Update, PSL, PMC servo restored
|
Sun Jan 26 13:47:00 2020, gautam, Update, PSL, PMC servo characterization   
|
Tue Jan 28 14:33:24 2020, gautam, Update, PSL, Inferred free-running frequency noise 
|
Thu Jan 23 18:52:31 2020, gautam, Update, PSL, PMC RFPD characterization
|
|
Message ID: 15163
Entry time: Tue Jan 28 14:33:24 2020
In reply to: 15156
|
Author: |
gautam |
Type: |
Update |
Category: |
PSL |
Subject: |
Inferred free-running frequency noise |
|
|
To conclude my PMC noise investigations: Attachment #1 shows the PMC noise inferred from the calibrations earlier in this thread and the fitted OLTF for the PMC loop. Attachment #2 compares the frequency noise (inferred from the error point of the PMC servo) when the IMC is locked / unlocked. I don't know what to make of the fact that the PMC suggests improvement from ~20 Hz onwards already - does this mean that the NPRO noise model is wrong by 1 order of magnitude at 30 Hz?
- The IMC was locked for the measurement shown in Attachment #1.
- The in-loop spectra of the error (at the I/F output of the mixer) and control (at TP3) signals were measured with the SR785.
- The control signal voltage monitors don't seem to work - neither the front panel LEMO nor the signals hooked up to the CDS system show me sensible shapes for the spectra between 1-3 Hz.
- To convert in loop to free-running, I multiplied the measured error (control) signal spectra by
( ), where L is the OLTF. THe control signal was pre-processed by multiplying by a pole at 11.3 Hz, corresponding to the LPF formed by the 63.3 kohm series resistor and the 225 nF PZT capacitance.
- The "NPRO noise model" curve is 10^4/f Hz/rtHz.
While I initially thought the 1/f^2 rise below ~100 Hz is attributable to the IMC cavity length fluctuations, I found that this profile is present even in the measurement with the PSL shutter closed. I am not embarking on a detailed PMC noise budgeting project for now. Note however that we are not shot noise limited anywhere in this measurement band.
- The measured TF (dots in Attachment #5) was fit with LISO (solid lines in Attachment #5) to allow inferring the out-of-loop servo noise by monitoring the in-loop noise (that plot to follow).
|
|
|
|