Wed Dec 11 15:01:57 2019, Jon, Summary, PSL, PMC cavity ringdown measurement
|
Thu Dec 12 19:20:43 2019, Yehonathan, Update, PSL, PMC cavity ringdown measurement  
|
Fri Dec 13 12:28:43 2019, Yehonathan, Update, PSL, PMC cavity ringdown measurement
|
Tue Dec 17 20:45:30 2019, rana, Update, PSL, PMC cavity ringdown measurement
|
Fri Dec 27 15:01:02 2019, Yehonathan, Update, PSL, PMC cavity ringdown measurement   
|
Tue Dec 31 03:03:02 2019, gautam, Update, PSL, PMC cavity ringdown measurement
|
Mon Dec 16 18:19:42 2019, shruti, Update, PSL, PMC cavity ringdown measurement : beat-note disruption
|
|
Message ID: 15098
Entry time: Mon Dec 16 18:19:42 2019
In reply to: 15096
|
Author: |
shruti |
Type: |
Update |
Category: |
PSL |
Subject: |
PMC cavity ringdown measurement : beat-note disruption |
|
|
I have removed the PD55 + ND filter attached to it (see Attachment) and placed it next to the oscilloscope, after disconnecting its output and power supply. The post is still in place.
I did see the beat after that.
Quote: |
{Yehonathan, Rana, Jon}
To check whether we laser is being shut fast enough for the ringdown measurement we put a PD55 in the path that leads to the beat note setup. The beam is picked off from the back steering mirror after AOM and before the PMC.
@Shruti the PD is now blocking the beam to your setup.
|
|
|
|