40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Dec 3 22:41:17 2019, shruti, Update, General, PLL for PM measurement  PM_measurement.jpeg
    Reply  Wed Dec 4 20:32:43 2019, gautam, Update, General, PLL for PM measurement  
    Reply  Mon Dec 9 21:22:46 2019, shruti, Update, General, PLL / PM measurement of Xend NPRO PZT 
       Reply  Tue Dec 17 20:08:09 2019, shruti, Update, General, PLL / PM measurement of Xend NPRO PZT 20191217.png
          Reply  Thu Jan 16 19:32:23 2020, shruti, Update, General, PLL / PM measurement of Xend NPRO PZT 
             Reply  Thu Jan 23 20:08:49 2020, shruti, Update, General, PLL / PM measurement of Xend NPRO PZT Image-1.jpg
                Reply  Tue Jan 28 19:40:15 2020, shruti, Update, General, PLL / PM measurement of Xend NPRO PZT PMTF.pdfBeatSpectrum.pdf
Message ID: 15069     Entry time: Tue Dec 3 22:41:17 2019     Reply to this: 15074     15088
Author: shruti 
Type: Update 
Category: General 
Subject: PLL for PM measurement  

I worked on the setup up for the phase modulation measurement of the X end NPRO PZT. A previous similar measurement can be found here (12077). The setup was assembled based on the schematic in Attachment1.

Mixer used: Level 7, Mini circuits ZP-3+
LPF: up to 1.9MHz


Cables exiting the PSL table:
1. LO (Marconi -> Mixer)
2. RF (PSL+X beat note -> Mixer) The cable for this was taken from the Beat Mouth (otherwise connected to the oscilloscope)
3. Ext modulator (SR560 -> Marconi)

The long cable labled 'X Green Beat' was used to connect to the PZT (from the network analyzer).

Observations: The beat note kept floating between 0 and ~100 MHz

The PLL part of the circuit was tested coarsely with the spectrum analyzer function of the Agilent, where the loop was seen to stabilize when the carrier frequency of the Marconi was close to the instantaneous beat frequency.

 

 