Tue Oct 8 18:35:09 2019, gautam, Update, LSC, Locking prep
|
Wed Oct 9 18:15:26 2019, gautam, Update, LSC, PRMI 3f locking
|
Wed Oct 9 22:02:58 2019, gautam, Update, LSC, REFL55 whitening issue
|
Thu Oct 10 01:12:56 2019, gautam, Update, LSC, Locking studies
|
Wed Oct 16 11:42:17 2019, gautam, Update, LSC, Poor separation of PRCL/MICH in 3f signals
|
Thu Oct 17 11:19:28 2019, gautam, Update, LSC, Locking activity last night 
|
Tue Oct 22 00:52:27 2019, gautam, Update, LSC, Locking updates
|
Tue Oct 29 01:24:45 2019, gautam, Update, LSC, More locking updates 
|
Tue Oct 29 15:13:19 2019, gautam, Update, LSC, More locking updates
|
Wed Oct 30 01:27:00 2019, gautam, Update, LSC, More locking updates
|
Wed Oct 30 11:53:41 2019, gautam, Update, LSC, MICH loop shape tuning 
|
Wed Oct 30 17:08:40 2019, gautam, Update, LSC, POP22 investigation
|
Mon Nov 4 15:29:47 2019, gautam, Update, LSC, POP signal path
|
Mon Nov 4 16:06:58 2019, gautam, Update, LSC, POP optical path    
|
Tue Nov 5 11:52:27 2019, gautam, Update, LSC, Locking notes
|
Wed Nov 6 02:08:48 2019, gautam, Update, LSC, Locking updates
|
Wed Nov 6 17:05:45 2019, gautam, Update, LSC, CARM calibration  
|
Wed Nov 6 17:45:34 2019, gautam, Update, LSC, ~
|
Mon Nov 18 21:04:38 2019, gautam, Update, LSC, Locking - some ideas
|
Wed Nov 20 01:07:18 2019, gautam, Update, LSC, Locking - progress
|
Wed Nov 20 12:14:17 2019, gautam, Update, LSC, Locking - progress
|
Wed Nov 20 17:20:24 2019, Yehonathan, Update, LSC, QPD Investigation
|
Wed Nov 20 17:52:00 2019, gautam, Update, LSC, QPD MEDM screen update  
|
Wed Nov 20 21:29:28 2019, gautam, Update, LSC, PRG ~13
|
Fri Nov 22 00:54:14 2019, gautam, Update, LSC, locking notes
|
Mon Nov 25 19:11:22 2019, gautam, Update, LSC, ALS noise re-look
|
Wed Nov 27 23:24:01 2019, gautam, Update, LSC, No shaking but no inspiration either
|
Sun Dec 1 13:38:48 2019, rana, Update, LSC, No shaking but no inspiration either
|
Tue Dec 3 00:03:57 2019, gautam, Update, LSC, Green ALS also shows elevated noise with high arm buildup
|
Wed Dec 4 19:54:27 2019, gautam, Update, LSC, A look to the past
|
Thu Dec 5 01:54:39 2019, gautam, Update, LSC, Partial CM board path engaged
|
Fri Dec 6 00:02:48 2019, gautam, Update, LSC, What is the correct way to set the 3f offsets?
|
Fri Nov 15 12:08:04 2019, gautam, Update, LSC, POPDC whitening board
|
Tue Oct 29 17:40:48 2019, gautam, Update, LSC, More locking updates 
|
|
Message ID: 15056
Entry time: Wed Nov 27 23:24:01 2019
In reply to: 15046
Reply to this: 15057
|
Author: |
gautam |
Type: |
Update |
Category: |
LSC |
Subject: |
No shaking but no inspiration either |
|
|
Summary:
I tried the following minor changes to the locking procedure to see if there were any differences in the ALS noise performance:
- Actuate DARM only on one ETM (tried both ETMX and ETMY)
- Enable MCL and PRC seismic feedforward
- DC couple the ITM Oplevs for better angular stability during the lock acquisition
None of these changes had any effect - the ALS noise still goes up with arm buildup.
I think a good way to determine if the problem is to do with the IR part of the new ALS system is to resurrect the green beat setup - I expect this to be less invasive than installing attenuators/beam dumps in front of the fiber couplers at the ends. We should at the very least recover the old ALS noise levels and we were able to lock the PRFPMI with that config. If the excess noise persists, we can rule out the problem being IR scatter into the beat-mouth fibers. Does this sound like a reasonable plan? |