Thu Nov 14 23:56:18 2019, rana, Update, LSC, off the bad Violin filters
|
Fri Nov 15 11:58:12 2019, gautam, Update, LSC, off the bad Violin filters 
|
Mon Mar 1 16:27:17 2021, gautam, Update, LSC, PRM violin filter excessive? 
|
Tue Mar 2 13:39:31 2021, rana, Update, LSC, PRM violin filter excessive?
|
|
Message ID: 15028
Entry time: Fri Nov 15 11:58:12 2019
In reply to: 15026
|
Author: |
gautam |
Type: |
Update |
Category: |
LSC |
Subject: |
off the bad Violin filters |
|
|
The clue was that the loop X arm POX loop looked to have low (<3dB)) gain margin around 600 Hz (and again at 700 Hz). Attachment #1 shows a comparison of the OLTF for this loop (measured using the IN1/IN2 method) before and after our change. We hypothesize that one of the violin filters that were turned off had non-unity DC gain, because I had to lower the loop gain by 20% after these turn-offs to have the same UGF. I updated the snap files called by the arm locking scripts.
I think I caught all the places where the FM settings are saved, but some locking scripts may still try and turn on some of these filters, so let's keep an eye on it.
Quote: |
We turned off many excessive violin mode bandstop filters in the LSC.
|
|
|
|
|
|