Message ID: 14984     Entry time: Tue Oct 22 15:32:15 2019     In reply to: 14981     Reply to this: 14986
Author: gautam 
Type: Update 
Category: ALS 
Subject: EX uPDH electronics checkout 

Summary:

The EX PDH setup had what I thought was insufficient phase and gain margins. So I lowered the gain a little - the price paid was that the suppression of laser frequency noise of the end laser was reduced. I actually think an intermediate gain setting (G=7) can give us ~35 degrees of phase margin, ~10dB gain margin, and lower residual unsuppressed AUX laser noise - to be confirmed by measurement later. See here for the last activity I did - how did the gain get increased? I can't find anything in the elog.

Attachment 1: uPDH_X_OLTFs.pdf  174 kB  Uploaded Tue Oct 22 18:22:27 2019  | Hide | Hide all
uPDH_X_OLTFs.pdf