40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Jun 25 15:32:16 2019, gautam, Update, IOO, PMC and IMC locked again, some MEDM maintenance PMCctrl.png
    Reply  Wed Jun 26 10:55:13 2019, aaron, Update, IOO, PMC and IMC locked again, some MEDM maintenance 
       Reply  Wed Jun 26 11:11:40 2019, Milind, Update, IOO, PMC and IMC locked again, some MEDM maintenance burtgooey.pdf
    Reply  Wed Jun 26 18:28:24 2019, rana, Update, IOO, PMC and IMC locked again, some MEDM maintenance 
       Reply  Sun Jun 30 22:21:07 2019, Milind, Update, IOO, PMC and IMC locked again, some MEDM maintenance 
          Reply  Sun Jun 30 23:52:09 2019, Koji, Update, IOO, PMC and IMC locked again, some MEDM maintenance 
             Reply  Mon Jul 1 11:02:05 2019, Milind, Update, IOO, PMC and IMC locked again, some MEDM maintenance 
          Reply  Tue Jul 9 10:37:42 2019, Milind, Update, IOO, keyed psl crate, unstick.py, pmc autolocker code- working  
             Reply  Mon Jul 15 14:53:40 2019, Milind, Update, IOO, keyed psl crate, unstick.py 
                Reply  Mon Jul 15 18:55:05 2019, gautam, Update, IOO, Megatron hard-rebooted 
             Reply  Wed Jul 24 12:24:43 2019, Milind, Update, IOO, unstick.py and ifotest 
Message ID: 14700     Entry time: Wed Jun 26 11:11:40 2019     In reply to: 14699
Author: Milind 
Type: Update 
Category: IOO 
Subject: PMC and IMC locked again, some MEDM maintenance 

After helping Aaron key the crate and do a burt restore, I realized that it would probably be best to record the steps that Koji showed me to do a burt restore as a reference for (anyone) the future

Commands (in terminal):

  1. burttoday: changes to the directory with snapshots for the day (/opt/rtcds/caltech/c1/burt/autoburt/today)
  2. burtgooey: opens a new window with several buttons of which "Restore" needs to be selected. This opens up a second window as shown in Attachment #1. Click on Snapshot files and navigate to the snapshot you wish to restore (these are present at /opt/rtcds/caltech/c1/burt/autoburt/snapshots) and select that. A green "OK" button indicates if the Restore can be performed without a hitch. Hit "Restore" to perform the burtrestore.

 

Also Gautam explained today that the sticky slider problem is a hardware issue. That it basically means that the signal (voltage output for instance) that you request from the medm screen is not what the hardware delivers. Twice now, we have got around that with a burtrestore. My understanding of a burt restore is that it is a restoration of values from a certain time to the EPICS channels. Therefore, I don't understand why a restoration at the software level should fix how the hardware responds? Why does this happen?

Attachment 1: burtgooey.pdf  17 kB  Uploaded Wed Jun 26 16:56:12 2019  | Hide | Hide all