40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Wed May 15 00:40:19 2019, gautam, Update, SUS, ETMY diagnosis plan
|
Wed May 15 10:57:57 2019, gautam, Update, SUS, EY chamber opened
|
Wed May 15 17:46:24 2019, gautam, Update, SUS, ETMY inspection 
|
Wed May 15 19:36:29 2019, Koji, Update, SUS, ETMY instepction
|
Thu May 16 13:07:14 2019, gautam, Update, SUS, First contact residue removal 
|
Thu May 16 23:31:55 2019, gautam, Update, SUS, ETMY suspension characterization 6x
|
Fri May 17 17:01:08 2019, gautam, Update, SUS, ETMY suspension characterization
|
Mon May 20 11:33:46 2019, gautam, Update, SUS, ITMY inspection
|
Mon May 20 17:12:57 2019, gautam, Update, SUS, ETMY LL adjustment
|
Mon May 20 22:06:07 2019, gautam, Update, SUS, ITMY also kicked
|
Tue May 21 00:15:21 2019, gautam, Update, SUS, Main objectives of vent achieved (?)   
|
Tue May 21 21:33:27 2019, gautam, Update, SUS, ETMY HR face cleaned
|
Wed May 22 11:53:50 2019, gautam, Update, SUS, ETMY EQ stops backed out
|
|
Message ID: 14629
Entry time: Tue May 21 21:33:27 2019
In reply to: 14628
Reply to this: 14630
|
Author: |
gautam |
Type: |
Update |
Category: |
SUS |
Subject: |
ETMY HR face cleaned |
|
|
[koji, gautam]
We executed this plan. Photos are here. Summary:
- Optic was EQ-stopped (face stops only)., with the OSEMs in situ. We tried to do this as evenly as possible to avoid any magnets getting stuck on OSEMs.
- We used the specially procured acetone from Chub to drag wipe the HR face. This was a definite improvement, we should always get the correct grade of solvents when we attempt cleaning optics.
- It was observed that drag-wiping did not really have the desired cleaning effect. So Koji went in with hemostat / lens tissue soaked in acetone and wiped the HR face. This improved the situation.
- Applied a layer of F.C. Waited for it to dry, and then peeled it off. Under the green flashlight, the optic still looks horrific - but we decided against further drag-wiping/first-contacting. If the loss is truly 50 ppm, this is totally not a show-stopper for now.
- Suspension cage was replaced. EQ stops were released. Bias voltages were adjusted to bring the Oplev spot back to the center of the QPD. Now a free-swinging data collection is ongoing...
The following optics were kicked:
ETMY
Tue May 21 22:58:18 PDT 2019
1242539916
So if nothing, we got to practise this new wiping technique with OSEMs in situ successfully.
Quote: |
- Prepare ETMY for first contact cleaning to remove the residual piece.
- Drag wipe the HR surface with dehydrated acetone
- Apply F.C. as usual, inspect the HR face after peeling for improvement if any.
- This will give us a chance to practise the F.C.ing with the optic EQ-stopped (moving cage etc).
|
|