Fri Apr 19 14:06:36 2019, gautam, Update, PSL, Innolight NPRO shutoff
|
Fri Apr 19 20:21:52 2019, gautam, Update, PSL, Innolight NPRO shutoff
|
Wed Apr 24 16:06:44 2019, gautam, Update, PSL, Innolight NPRO shutoff
|
Thu Apr 25 01:03:29 2019, gautam, Update, PSL, MC trans is ~1000 cts (~7%) lower than usual
|
Thu Apr 25 17:31:56 2019, gautam, Update, PSL, Innolight NPRO shutoff 
|
Mon Apr 29 16:25:22 2019, gautam, Update, PSL, PSL turned on again
|
Fri May 3 12:51:58 2019, gautam, Update, PSL, PSL turned on again
|
Mon May 6 10:51:43 2019, gautam, Update, PSL, PSL turned off again
|
Wed May 8 19:04:20 2019, rana, Update, PSL, PSL turned on again
|
Thu May 9 19:50:04 2019, gautam, Update, PSL, PSL turned off again
|
Mon May 13 10:45:38 2019, gautam, Update, PSL, PSL turned ON again
|
Wed May 15 10:56:47 2019, gautam, Update, PSL, PSL turned ON again
|
|
Message ID: 14595
Entry time: Mon May 6 10:51:43 2019
In reply to: 14593
|
Author: |
gautam |
Type: |
Update |
Category: |
PSL |
Subject: |
PSL turned off again |
|
|
As we have seen in the last few weeks, the laser turned itself off after a few hours of running. So bypassing the lab interlock system / reverting laser crystal temperature to the value from Innolight's test datasheet did not fix the problem.
I do not understand why the "Interlock" and "TGUARD" channels come revert to their values when the laser was lasing a few minutes after the shutoff. Is this just an artefact of the way the diagnostics is set up, or is this telling us something about what is causing the shutoff? |
|
|