Tue Apr 2 18:40:01 2019, gautam, Update, VAC, Vac failure
|
Wed Apr 3 09:07:46 2019, gautam, Update, VAC, Vac failure
|
Wed Apr 3 10:42:36 2019, gautam, Update, VAC, TP2 forepump replaced
|
Wed Apr 3 16:17:17 2019, Jon, Update, VAC, TP2 forepump replaced
|
Wed Apr 3 18:35:54 2019, gautam, Update, VAC, PSL shutter re-opened
|
Fri Apr 5 01:10:18 2019, gautam, Update, VAC, TP3 forepump is also noisy
|
Tue Apr 16 22:06:51 2019, gautam, Update, VAC, Vac interlock tripped again 
|
Wed Apr 17 18:12:06 2019, gautam, Update, VAC, Vac interlock tripped again 
|
|
Message ID: 14511
Entry time: Wed Apr 3 09:07:46 2019
In reply to: 14509
Reply to this: 14512
|
Author: |
gautam |
Type: |
Update |
Category: |
VAC |
Subject: |
Vac failure |
|
|
Overnight pressure trends don't suggest anything went awry after the initial interlock trip. Some watchdog script that monitors vacuum pressure and closes the PSL shutter in the event of pressure exceeding some threshold needs to be implemented. Another pending task is to make sure that backup disk for c1vac actually is bootable and is a plug-and-play replacement. |
|
|