Message ID: 14436     Entry time: Tue Feb 5 19:30:14 2019     In reply to: 14434     Reply to this: 14438
Author: gautam 
Type: Update 
Category: VAC 
Subject: Main volume at 20 uTorr 

Pumpdown looks healthy, so I'm leaving the TPs on overnight. At some point, we should probably get the RGA going again. I don't know that we have a "reference" RGA trace that we can compare the scan to, should check with Steve. The high power (1 W) beam has not yet been sent into the vacuum, we should probably add the interlock condition that shuts off the PSL shutter before that.

Attachment 1: PD83.png  34 kB  | Hide | Hide all
PD83.png