Message ID: 14405     Entry time: Fri Jan 18 15:34:37 2019
Author: gautam 
Type: Update 
Category: Thermal Compensation 
Subject: Elliptical reflector part number 

Nobody documented this, but here is the part number with mechanical drawings of the elliptical reflector installed at EY: Optiforms E180. Heater is from Induceramics, but I can't find the exact part which matches the dimensions of the heater we have (diameter = 3.8mm, length = 30mm), perhaps it's a custom part?

The geometry dictates that if we want the heater to be at one focus and the ETM to be at the other, the separation has to be 7.1 inches. It certainly wasn't arranged this way before. It seems unrealistic to do this without clipping the main beam, I propose we leave sufficient clearane between the main beam and the reflector, and accept the reduced heating efficiency. 

Thanks to Steve for digging this up from his secret stash.