Entry  Thu Dec 20 17:17:41 2018, gautam, Update, CDS, Logging of new Vacuum channels 
    Reply  Fri Dec 21 11:11:51 2018, gautam, Update, CDS, Logging of new Vacuum channels 
Message ID: 14374     Entry time: Thu Dec 20 17:17:41 2018     Reply to this: 14376
Author: gautam 
Type: Update 
Category: CDS 
Subject: Logging of new Vacuum channels 

Added the following channels to C0EDCU.ini:

[C1:Vac-P1b_pressure]
units=torr
[C1:Vac-PRP_pressure]
units=torr
[C1:Vac-PTP2_pressure]
units=torr
[C1:Vac-PTP3_pressure]
units=torr
[C1:Vac-TP2_rot]
units=kRPM
[C1:Vac-TP3_rot]
units=kRPM

Also modified the old P1 channel to

[C1:Vac-P1a_pressure]
units=torr

Unfortunately, we realized too late that we don't have these channels in the frames, so we don't have the data from this test pumpdown logged, but we will have future stuff. I say we should also log diagnostics from the pumps, such as temperature, current etc. After making the changes, I restarted the daqd processes.


Things to add to ASA wiki page once the wiki comes back online:

  1. What is the safe way to clean the cryo pump if we want to use it again?
  2. What are safe conditions to turn the RGA on?