40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Sep 6 01:09:11 2017, johannes, Update, Computer Scripts / Programs, New set of loss measurements XARM_20170905.pdfYARM_20170905.pdf
    Reply  Mon Sep 11 12:56:40 2017, johannes, Update, Computer Scripts / Programs, lossmap attempts 
       Reply  Mon Sep 11 16:46:09 2017, Steve, Update, PEM, earthquakes M5.4eq.png
          Reply  Tue Sep 19 14:00:41 2017, Steve, Update, PEM, earthquakes 7.1MX5.3J.png
             Reply  Fri Oct 27 09:34:20 2017, Steve, Update, PEM, earthquakes recentEQs.png
                Reply  Fri Nov 10 13:46:26 2017, Steve, Update, PEM, M3.1 local earthquake 3.1M_local_EQ.png
                   Reply  Mon Apr 9 08:39:39 2018, Steve, Update, PEM, M5.3 eq Souther CA M5.3_Santa_Cruz_Is.CA.pngafter_M5.3.pngM5.3vac.png
                Reply  Thu Jan 25 08:27:25 2018, Steve, Update, PEM, earthquakes M4_local_eq.pngM4.pngM5.8Ferndale_CA.png
       Reply  Mon Sep 25 17:56:33 2017, johannes, Update, Computer Scripts / Programs, transmitted power during lossmap 
    Reply  Thu Oct 11 13:40:51 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
       Reply  Fri Oct 12 12:29:34 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
          Reply  Fri Oct 12 20:20:29 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
             Reply  Sat Oct 13 20:11:10 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
                Reply  Mon Oct 15 10:32:13 2018, yuki, Update, Computer Scripts / Programs, loss measurements XARM_20181015_1500.pdfYARM_20181015_1500.pdf
                   Reply  Tue Oct 16 00:44:29 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
                   Reply  Fri Nov 2 19:25:16 2018, gautam, Update, Computer Scripts / Programs, loss measurements 
    Reply  Mon Oct 15 12:52:54 2018, yuki, Update, Computer Scripts / Programs, additional comments 
       Reply  Fri Nov 2 16:42:31 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents 
          Reply  Mon Nov 5 13:52:18 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents 
             Reply  Tue Nov 6 10:19:26 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents 
                Reply  Tue Nov 6 15:23:48 2018, gautam, Update, IOO, IMC problematic MCwonky.png
                   Reply  Tue Nov 6 19:02:35 2018, aaron, Update, IOO, IMC problematic 181106_MCTRANS.jpg
                      Reply  Sat Nov 10 17:40:00 2018, aaron, Update, IOO, IMC problematic 
                         Reply  Mon Nov 12 13:53:20 2018, rana, Update, IOO, loss measurement: oscope vs CDS DAQ 
                         Reply  Thu Nov 15 10:21:07 2018, aaron, Update, IOO, IMC problematic 181115_chansDown.pngPD_noise.png
                            Reply  Fri Nov 16 10:53:07 2018, aaron, Update, IOO, IMC problematic Screenshot_from_2018-11-16_19-22-34.png
                               Reply  Sat Nov 17 18:59:01 2018, aaron, Update, IOO, IMC problematic 
                Reply  Wed Nov 7 05:16:16 2018, yuki, Update, Computer Scripts / Programs, arm loss measuremenents 
Message ID: 14274     Entry time: Tue Nov 6 10:19:26 2018     In reply to: 14270     Reply to this: 14275   14280
Author: aaron 
Type: Update 
Category: Computer Scripts / Programs 
Subject: arm loss measuremenents 

I'm checking out the data this morning, running armloss_AS_calc.py using the parameters Yuki used here.

I made the following changes to scripts (measurement script and calculator script)

  • Included the 'hour' of the run in the armloss_dcrefl_* script. This way, we can run more than once a day without overwriting data.
  • Changed the calculator script to loop over all iterations of locked/misaligned states, and calculate the loss for adjacent measurements.
    • That is, the measurement script will make a measurement with the arm locked, then with it misaligned, and repeat that N times
    • The calculator now finds the loss for the nth iteration using *_n_locked and *_n_misaligned, and finds N separate loss measurements
    • The dark signal is also computed N times, though all of the dark measurements are made before running the arm scripts, so they could be all integrated together.
    • All of these are saved in the same directory that the data was grabbed from.

I repeated the 'dark' measurements, because I need 20 files to run the script and the measurements before had the window on the scope set larger than the integration time in the script, so it was padded with bad values that were influencing the calculation.

On running the script again, I'm getting negative values for the loss. I removed the beamstops from the PDs, and re-centered the beams on the PDs to repeat the YARM measurements.