40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Sep 6 01:09:11 2017, johannes, Update, Computer Scripts / Programs, New set of loss measurements XARM_20170905.pdfYARM_20170905.pdf
    Reply  Mon Sep 11 12:56:40 2017, johannes, Update, Computer Scripts / Programs, lossmap attempts 
       Reply  Mon Sep 11 16:46:09 2017, Steve, Update, PEM, earthquakes M5.4eq.png
          Reply  Tue Sep 19 14:00:41 2017, Steve, Update, PEM, earthquakes 7.1MX5.3J.png
             Reply  Fri Oct 27 09:34:20 2017, Steve, Update, PEM, earthquakes recentEQs.png
                Reply  Fri Nov 10 13:46:26 2017, Steve, Update, PEM, M3.1 local earthquake 3.1M_local_EQ.png
                   Reply  Mon Apr 9 08:39:39 2018, Steve, Update, PEM, M5.3 eq Souther CA M5.3_Santa_Cruz_Is.CA.pngafter_M5.3.pngM5.3vac.png
                Reply  Thu Jan 25 08:27:25 2018, Steve, Update, PEM, earthquakes M4_local_eq.pngM4.pngM5.8Ferndale_CA.png
       Reply  Mon Sep 25 17:56:33 2017, johannes, Update, Computer Scripts / Programs, transmitted power during lossmap 
    Reply  Thu Oct 11 13:40:51 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
       Reply  Fri Oct 12 12:29:34 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
          Reply  Fri Oct 12 20:20:29 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
             Reply  Sat Oct 13 20:11:10 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
                Reply  Mon Oct 15 10:32:13 2018, yuki, Update, Computer Scripts / Programs, loss measurements XARM_20181015_1500.pdfYARM_20181015_1500.pdf
                   Reply  Tue Oct 16 00:44:29 2018, yuki, Update, Computer Scripts / Programs, loss measurements 
                   Reply  Fri Nov 2 19:25:16 2018, gautam, Update, Computer Scripts / Programs, loss measurements 
    Reply  Mon Oct 15 12:52:54 2018, yuki, Update, Computer Scripts / Programs, additional comments 
       Reply  Fri Nov 2 16:42:31 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents 
          Reply  Mon Nov 5 13:52:18 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents 
             Reply  Tue Nov 6 10:19:26 2018, aaron, Update, Computer Scripts / Programs, arm loss measuremenents 
                Reply  Tue Nov 6 15:23:48 2018, gautam, Update, IOO, IMC problematic MCwonky.png
                   Reply  Tue Nov 6 19:02:35 2018, aaron, Update, IOO, IMC problematic 181106_MCTRANS.jpg
                      Reply  Sat Nov 10 17:40:00 2018, aaron, Update, IOO, IMC problematic 
                         Reply  Mon Nov 12 13:53:20 2018, rana, Update, IOO, loss measurement: oscope vs CDS DAQ 
                         Reply  Thu Nov 15 10:21:07 2018, aaron, Update, IOO, IMC problematic 181115_chansDown.pngPD_noise.png
                            Reply  Fri Nov 16 10:53:07 2018, aaron, Update, IOO, IMC problematic Screenshot_from_2018-11-16_19-22-34.png
                               Reply  Sat Nov 17 18:59:01 2018, aaron, Update, IOO, IMC problematic 
                Reply  Wed Nov 7 05:16:16 2018, yuki, Update, Computer Scripts / Programs, arm loss measuremenents 
Message ID: 14268     Entry time: Fri Nov 2 16:42:31 2018     In reply to: 14255     Reply to this: 14270
Author: aaron 
Type: Update 
Category: Computer Scripts / Programs 
Subject: arm loss measuremenents 

I'm continuing the arm loss measurements Yuki was making. I'm first familiarizing myself with the procedures for the measurement Johannes describes.

I'm not very familiar with the medm screens, so I'm just kind of poking around and checking with Gautam. I do the following:

  1. Turned Xarm ASS dither on, then off.
  2. Turned X and Y ALS on, then off shortly after
    1. Realizing I needed some guidance, I found this page on lock acquisition on the wiki
    2. Gautam showed me how to align/lock the IFO so I could take some notes, and we locked the Y arm, misaligned X.
  3. I put the PD back in the AS beam path to get the ASDC signal, and approximately centered the beam. This PD is on channel 1 of the scope, which is at 192.168.113.24.
  4. I centered the beam onto the MC2 PD that Yuki had installed. This PD is on channel 2 of the scope.
    1. Both scope channels are set to 1V scale (I also had tried 500mV, and it didn't seem to make a difference) and 10s time axis spacing (maximum integration time, since we're looking for a DC effect. Is this what we want?)
    2. The impedance for both channels is 1MOhm.
  5. I ran the script to start the loss measurement on the Y arm.
    1. python2 armloss_dcrefl_asdcpd_scope.py 192.168.113.24 1 2 5 YARM
    2. I'm reading ~15 (au?) for the MC channel and ~5% of that out the AS, which seems to make sense to me and looked to be about what Yuki the ratios when I checked the log files. However, I'm a bit confused by the normalization, because the maximum output of the MC PD is 10V, and indeed the scope's display is reading under 10V.

I've left the script running.