I was preparing for the aLIGO EOM measuement to be carried out tomorrow afternoon.

I did a few modifications to the PLL setup.

  • The freq mixier in the PLL setup was replaced with ZP3 (level 7) from ZAD-6
  • The PLL gain was reduced from 3.10 to 2.80 to prevent servo oscillation
  • The main PSL marconi is connected to the PLL mixier and providing fixed 200MHz 8dBm.
  • The main PSL modulation is off.

Tomorrow I am going to modulate the EOM with the AUX Marconi via an amplifier (probably)

Automated scripts (AGinit.py and AGmeas.py) are in /users/koji/scripts

I will revert the setup once the measurement is done tomorrow.