Message ID: 14096     Entry time: Sat Jul 21 14:03:19 2018     In reply to: 14094
Author: Koji 
Type: Summary 
Category: Thermal Compensation 
Subject: Y arm locking 

Ah. With MC2 feedback, we have about 3 times smaller "optical gain" for the ASS A2L. We have same dither, same actuator, but we need only 1/3 actuation of the MC2 compared to the ETMY case.
We had to reduce the ASS spot servo from 1 to 0.3 to make is stable, so this means that the ASS is really merginally stable.