[Annalisa, Terra, Koji, Gautam]
Summary: We find a configuration for arm scans which significantly reduces phase noise. We run several arm scans and we were able to resolve several HOM peaks; analysis to come.
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
As first, we made a measurement with the already established setup and, as Jon already pointed out, we found lots of phase noise. We hypothesized that it could either come from the PLL or from the motion of the optics between the AUX injection point (AS port) and the Y arm.
- We first characterized the PLL loop phase noise by comparing the beat signal against the Agilent reference signal, and we found that the beat had lots of phase noise with respect to the reference. Decreasing the PLL gain, we got rid of the phase noise in the beat signal.
- Next, for the optical path length induced phase noise, we took the transfer function between TransMon and REFL signal rather than TransMon and Agilent reference signal. This takes advatage of the fact that the TransMon and REFL both see optical path length phase noise, which therefore gets canceled out in the transfer function.
In this configuration, we were able to do arm scans where the phase variation at each peak was pretty clear and well defined. We took several 10MHz scan, we also zoomed around some specific HOM peak, and we were able to resolve some frequency split.
We add some pictures of the setup and of the scan.